Weinmann OXYWAY Fix I Service Manual page 68

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fix I, OXYWAY Fix III
114
113
113
114
113
103
68
EN
OXYWAY
Back
13.6 Replacing the O-ring in the
connecting bolt
Required tools
• Watchmaker's screwdriver 409
• Cloth
13.6.1 Removing the O-ring from the connecting bolt
1. Use the watchmaker's screwdriver 409 to carefully lift up the
13.6.2 Fitting the O-ring in the connecting bolt
1. If the groove in the connecting bolt 113 is soiled:
2. Insert the O-ring 114 into the groove in the connecting bolt 113
13.7 Replacing the disposable filter
Required tools
Allen key size 5
13.7.1 Removing the disposable filter
1. Use the Allen key to loosen the disposable filter 103 in
2. Remove the disposable filter 103 from the connecting bolt 113.
Forward
O-ring 114 and remove it from the groove in the connecting
bolt 113.
When doing so, note: The groove must not be damaged.
Clean the groove in the connecting bolt 113 with a dry cloth.
by hand.
the connecting bolt 113.