Repairs to OXYWAY Fix I, OXYWAY Fix III
114
113
113
114
113
103
68
EN
OXYWAY
Back
13.6 Replacing the O-ring in the
connecting bolt
Required tools
• Watchmaker's screwdriver 409
• Cloth
13.6.1 Removing the O-ring from the connecting bolt
1. Use the watchmaker's screwdriver 409 to carefully lift up the
13.6.2 Fitting the O-ring in the connecting bolt
1. If the groove in the connecting bolt 113 is soiled:
2. Insert the O-ring 114 into the groove in the connecting bolt 113
13.7 Replacing the disposable filter
Required tools
Allen key size 5
13.7.1 Removing the disposable filter
1. Use the Allen key to loosen the disposable filter 103 in
2. Remove the disposable filter 103 from the connecting bolt 113.
Forward
O-ring 114 and remove it from the groove in the connecting
bolt 113.
When doing so, note: The groove must not be damaged.
Clean the groove in the connecting bolt 113 with a dry cloth.
by hand.
the connecting bolt 113.
Contents
Previous page
Next page