Repairs to OXYWAY Fine I
404
409
404
184
EN
OXYWAY
Back
5. Push the counter cap 404 over the safety valve and the housing.
6. Place the torque wrench onto the square of the counter cap 404.
7. Use the torque wrench to loosen the housing from the connecting
271
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 271 by hand and
10.Use the watchmaker's screwdriver 409 to remove the rib filter 256
256
271
16.12.2 Fitting the rib filter
1. Insert the rib filter 256 into the connecting bolt 271.
256
271
2. Screw the housing onto the connecting bolt 271 by hand.
3. Push the counter cap 404 over the safety valve and the housing.
4. Place the torque wrench onto the square of the counter cap 404.
5. Use the torque wrench to tightly screw the housing onto the
271
6. Remove the torque wrench and counter cap 404.
7. Loosen the hand-screwed connection 273 at the top part of the
8. Unscrew the hand-screwed connection 273 from the bottom part
9. Remove the hand-screwed connection 273 of the pressure
10.Remove the top part of the counter tool 400 from the bottom
Forward
bolt 271.
then remove.
from the connecting bolt 271.
connecting bolt 271.
counter tool 400.
of the counter tool 400 by hand.
reducer from the counter tool 400.
part of the counter tool 400.