17.6 Replacing the sieve and nozzle plate
in the inhalation outlet
Required tools
• Watchmaker's screwdriver 409
• Tweezers
17.6.1 Removing the sieve and nozzle plate from the
inhalation outlet
Requirement
The pressure screw 302 has been removed from the inhalation
outlet 301
inhalation outlet", page
1. Use the watchmaker's screwdriver 409 to carefully remove the
sieve 303 and sealing disks 304 from the inhalation outlet 301.
17.6.2 Fitting the sieve and sealing disks in the
inhalation outlet
1. Use tweezers to insert the sealing disk 304 in the inhalation
outlet 301.
2. Use tweezers to insert the sieve 303 in the inhalation outlet 301.
3. Use tweezers to insert the sealing disk 304 in the inhalation
outlet 301.
4. Fit the pressure screw 302 in the inhalation outlet 301
(see "17.5.2 Fitting the pressure screw in the inhalation outlet",
page
17.7 Replacing the inhalation outlet
Required tools
• Cover clamping device 401
• Counter tool with fastening handle 400
• Open-end wrench A/F 15
• Vise
• Brass brush
• Loctite 413
• Torque wrench 8 Nm ± 2 Nm
• Thread tap G 1/8
17.7.1 Removing the inhalation outlet
Requirement
• The pressure screw 302 has been removed from the inhalation
outlet 301
inhalation outlet", page
• The sieve 303 and the sealing disks 304 have been removed from
the inhalation outlet 301
nozzle plate from the inhalation outlet", page
Contents
(see "17.5.1 Removing the pressure screw from the
208).
208).
(see "17.5.1 Removing the pressure screw from the
208).
(see "17.6.1 Removing the sieve and
Previous page
Repairs to OXYWAY Click
209).
OXYWAY
EN
209
Next page

Advertisement

Table of Contents
loading

Table of Contents