Pressure reducer
Table of Contents

Advertisement

305
Back
Forward
17.11.2 Fitting the threaded pin in the multistep flow
selector
1. Insert the threaded pin 306 into the multistep flow selector 305 and
306
screw in using the Allen key.
When doing so, note: The threaded pin 306 must not be screwed
into the hole for the orifice plate assembly 325.
2. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
3. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
17.12 Replacing the multistep flow selector
17.12.1 Removing the multistep flow selector
Requirement
• The multistep flow selector assembly 323 has been removed
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
• The multistep flow selector assembly 323 has been disassembled
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
• The threaded pin 306 has been removed from the multistep flow
selector 305
multistep flow selector", page
1. Dispose of the multistep flow selector 305.
17.12.2 Fitting the multistep flow selector
1. Pick up the multistep flow selector 305.
2. Fit the threaded pin 306 in the multistep flow selector 305
(see "17.11.2 Fitting the threaded pin in the multistep flow
selector", page
3. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
4. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).