Repairs to OXYWAY Click
307
314
313
305
310
309
315
214
EN
OXYWAY
Back
6. Align the orifice plate assembly 325 in the cover assembly 324.
7. Carefully place the cover assembly 324 and the orifice plate
8. Insert the needle 315.
9. Insert the sliding ring 310.
10.Insert the springs 311.
11.Place the balls 309 on the springs 311.
12.Place the multistep flow selector 305 on the cover assembly 324
311
324
13.Fix the multistep flow selector 305 in place using the clamping
417
305
14.Use the Allen key to tightly screw in the threaded pin 306 on the
15.Loosen the fastening on the clamping device for cover
16.Remove the multistep flow selector assembly 323 from the
17.Check the alignment of the orifice plate assembly 325 in the
Forward
When doing so, note: The bores in the multistep flow selector 307,
the bore in the orifice plate 314, and the grooved pin in the holder
for the orifice plate 313 must be in a straight line.
assembly 325 back onto the clamping device for cover
assembly 417.
When doing so, note: The position of the orifice plate assembly 325
must not change.
with the needle 315, sliding ring 310, springs 311 and balls 309.
When doing so, note: The "0" on the multistep flow selector 305
must be at the level of the needle 315.
device for cover assembly 417.
multistep flow selector 305.
assembly 417.
clamping device for cover assembly 417.
multistep flow selector assembly 323.