Weinmann OXYWAY Fix I Service Manual page 66

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fix I, OXYWAY Fix III
66
EN
OXYWAY
Back
13.3 Replacing a label
Required tools
• Knife
• Spirit or Label remover
13.3.1 Removing a label
1. Remove the label 116 from the pressure reducer.
116
2. Use spirit or label remover to remove any residue from the pressure
3. Dispose of the label 116.
13.3.2 Attaching a label
1. Pull the protective film off the back of the label 116.
2. Stick the label 116 on the pressure reducer at the previously
116
13.4 Replacing the pressure screw in the
pressure outlet/inhalation outlet
Required tools
Slot-head screwdriver
13.4.1 Removing the pressure screw from the pressure
outlet/inhalation outlet
1. Use the slot-head screwdriver to loosen the pressure screw 108
2. Remove the pressure screw 108 from the pressure outlet 105/
105/106
108
Forward
reducer.
When doing so, note: Do not touch the adhesive surface of
the label 116.
used position.
When doing so, note: The adhesive surface must be free from
dust and grease.
in the pressure outlet 105/inhalation outlet 106.
inhalation outlet 106.