10.Remove the top part of the counter tool 400 from the bottom
part of the counter tool 400.
11.Loosen the bottom part of the counter tool 400 from the vise
and remove.
12.Fit the O-ring 114 in the connecting bolt 113
(see "13.6.2 Fitting the O-ring in the connecting bolt", page
13.11 Replacing the pressure outlet/
inhalation outlet
Required tools
• Counter tool with fastening handle 400
• Vise
• Loctite 413
• Torque wrench 25 Nm± 2 Nm
• Open-end wrench A/F 15
• Allen key size 9
• Brass brush
• Thread tap G 1/8
13.11.1 Removing the pressure outlet/inhalation outlet
Requirement
• The O-ring 114 has been removed from the connecting bolt 113
(see "13.6.1 Removing the O-ring from the connecting bolt",
page
• The pressure screw 108 has been removed from the pressure
outlet 105/inhalation outlet 106
pressure screw from the pressure outlet/inhalation outlet",
page
• The sealing disks 109 and the nozzle plate 107 have been removed
from the inhalation outlet 106
disks and nozzle plate from the inhalation outlet", page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 115 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 115 onto the top part
of the counter tool 400.
5. Use the open-end wrench/Allen key to loosen the pressure
outlet 105/inhalation outlet 106 in the housing.
Contents
Repairs to OXYWAY Fix I, OXYWAY Fix III
68).