16.26.2 Fitting the flow indicator
1. If there is dirt inside the housing: Remove the dirt with a dry,
lint-free cloth.
2. Place the sealing disk 253 into the threaded hole in the flow
251
indicator by hand.
3. Tighten the flow indicator 251 with the 23 Nm torque wrench.
4. Check the position of the flow indicator 251.
When doing so, note: The scale on the flow indicator 251 must be
facing forwards.
5. If the scale on the flow indicator 251 is not facing forwards:
• Use the open-end wrench to loosen the flow indicator 251 and
then remove it.
• Place the additional sealing disk 252 or 253 into the threaded
hole in the flow indicator 251 by hand.
• Tighten the flow indicator 251 with the 23 Nm torque wrench.
• Check the position of the flow indicator 251.
When doing so, note: The scale on the flow indicator must be
facing forwards.
• If necessary: Re-adjust the flow indicator 251 with the 28 Nm
torque wrench.
6. Loosen the hand-screwed connection 273 at the top part of the
counter tool 400.
7. Unscrew the hand-screwed connection 273 from the bottom part
of the counter tool 400 by hand.
8. Remove the hand-screwed connection 273 of the pressure reducer
from the counter tool 400.
9. Remove the top part of the counter tool 400 from the bottom part
of the counter tool 400.
10.Loosen the bottom part of the counter tool 400 from the vise and
remove.
11.Fit the pressure gauge cap as protection against impact and
falling 254
protection against impact and falling", page
12.Fit the O-ring 272 in the connecting bolt 271
(see "16.8.2 Fitting the O-ring in the connecting bolt", page
Contents
(see "16.7.2 Fitting the pressure gauge cap as
Previous page
Repairs to OXYWAY Fine I
179).
180).