14.Read off the flow values at the testing device 512 and enter in
15.Remove the special kit 500 from the testing device 512.
16.Close the valve on the oxygen cylinder.
17.Remove the adapter 507 from the inhalation outlet 205.
18.Remove the cap 503 from the pressure outlet 209.
19.Remove the hand-screwed connection 230 from the valve of
9.9 Checking the accessories
1. Also check the accessories according to the following table:
Requirement
• The oxygen cylinder shows no signs of external damage.
• The valve of the oxygen cylinder is fully functional.
• The pressure reducer on the oxygen cylinder is fully functional.
• The oxygen cylinder has a valid test certificate.
• The pressure hose shows no signs of external damage.
• The O-ring in the pressure hose shows no signs of external damage.
• The pressure hose label is present and legible.
• The connection nozzle shows no signs of external damage.
• The O-ring in the connection nozzle shows no signs of external damage.
The instructions for use are present.
Forward
the test report.
Requirement:
The flow value is 0 l/min ± 0.15 l/min.
the oxygen cylinder.