Weinmann OXYWAY Fix I Service Manual page 213

Pressure reducer
Table of Contents

Advertisement

324
Back
Forward
7. Use the Allen key to push the orifice plate assembly 325 into the
inside of the housing cover 308.
325
308
8. Remove the orifice plate assembly 325.
9. If necessary: Disassemble the orifice plate assembly 325
(see "17.10.1 Disassembling the orifice plate assembly",
page
215).
10.If necessary: Disassemble the cover assembly 324
(see "17.17.1 Disassembling the cover assembly", page
17.9.2 Assembling the multistep flow selector
assembly
1. If necessary: Assemble the cover assembly 324
(see "17.17.2 Assembling the cover assembly", page
2. If necessary: Assemble the orifice plate assembly 325
(see "17.10.2 Assembling the orifice plate assembly", page
3. Pick up the orifice plate assembly 325 and place it onto the
clamping device for cover assembly 417.
325
417
4. Place the cover assembly 324 onto the orifice plate assembly 325.
325
5. Remove the cover assembly 324 from the clamping device for cover
assembly 417 together with the orifice plate assembly 325.
Contents
Repairs to OXYWAY Click
OXYWAY
Previous page
220).
222).
215).
EN
213
Next page