14.Remove the cover assembly 180 from the cover clamping
15.Screw the cover assembly 180 onto the housing by hand.
16.Place the housing with the cover assembly 180 into the cover
17.Tighten the cover clamping device 401 in the vise.
18.Push the counter cap 404 over the safety valve and the housing.
19.Place the 8 Nm torque wrench onto the square of the counter
180
20.Use the 8 Nm torque wrench to tightly screw the housing onto
21.Remove the 8 Nm torque wrench and counter cap 404.
22.Remove the housing with the cover assembly 180 from the cover
23.Remove the cover clamping device 401 from the vise.
24.Clamp the bottom part of the counter tool 400 in the vise.
25.Put the top part of the counter tool 400 onto the bottom part of
26.Screw the hand-screwed connection 177 of the pressure reducer
27.Tightly screw the hand-screwed connection 177 onto the top part
28.Move the multistep flow selector 161 with inhalation outlet 155
29.Turn the multistep flow selector 161 with inhalation outlet 155
30.Place the special gripping pliers 415 on the union nut 170 and hold
415
31.Place the 16 Nm torque wrench with socket attachment onto the
32.Tighten the multistep flow selector 161 with inhalation outlet 155
155
33.Loosen the hand-screwed connection 177 at the top part of the
34.Unscrew the hand-screwed connection 177 from the bottom
35.Remove the housing with the cover assembly 180 from
Forward
device 401.
clamping device 401.
When doing so, note: The cover clamping device 401 must clamp
onto the housing cover 162.
cap 404.
the cover assembly 180.
clamping device 401.
the counter tool 400.
onto the bottom part of the counter tool 400 by hand.
of the counter tool 400.
to the desired position by hand.
around 30° counter-clockwise by hand in order to reach the desired
position after tightly screwing on.
in place.
inhalation outlet 155.
using the special gripping pliers 415 and the torque wrench.
When doing so, note: The torque wrench must activate in the
desired position at 16 Nm.
counter tool 400.
part of the counter tool 400 by hand.
the counter tool 400.