Weinmann OXYWAY Fix I Service Manual page 141

Pressure reducer
Table of Contents

Advertisement

Back
Forward
15.14 Replacing the pressure outlet
Required tools
• Counter tool with fastening handle 400
• Vise
• Loctite 413
• Torque wrench 25 Nm ± 1 Nm
• Open-end wrench A/F 15
• Brass brush
• Thread tap G 1/8
15.14.1 Removing the pressure outlet
Requirement
• The O-ring 229 has been removed from the connecting bolt 228
(see "15.8.1 Removing the O-ring from the connecting bolt",
page
• The pressure screw 210 has been removed from the pressure
outlet 209
pressure outlet", page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 230 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 230 onto the top
part of the counter tool 400.
5. Use the open-end wench to loosen the pressure outlet 209
from the housing.
6. Remove the pressure outlet 209 from the housing.
209
7. Remove any adhesive residue from the threads on the pressure
reducer housing.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
Contents
134).
(see "15.7.1 Removing the pressure screw from the
133).
Previous page
Repairs to OXYWAY Fast II
OXYWAY
EN
141
Next page

Advertisement