Weinmann OXYWAY Fix I Service Manual page 212

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Click
212
EN
OXYWAY
Back
5. Use the torque wrench to tighten the housing 320 with O
322
320
323
6. Loosen the vise.
7. Remove the pressure reducer from the cover clamping device 401.
8. Remove the cover clamping device 401 from the vise.
17.9 Disassembling/assembling the
multistep flow selector assembly
Required tools
• Allen key size 2
• Clamping device for cover assembly 417
17.9.1 Disassembling the multistep flow selector
assembly
Requirement
• The multistep flow selector assembly 323 has been removed
1. Use the Allen key to slightly loosen the threaded pin 306 on the
WE
Risk of injury from balls and springs springing out!
There are balls and springs underneath the multistep flow selector
that are under tension. They can spring out at the user or bystanders
and cause injury.
 Carefully remove the multistep flow selector.
2. Carefully remove the multistep flow selector 305.
305
The balls and sliding ring on the bottom of the multistep flow selector
may stick.
3. Remove the balls 309.
4. Remove the springs 311.
5. Remove the sliding ring 310.
6. Remove the needle 315.
Forward
sampling point 322 on the multistep flow selector assembly 323.
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
multistep flow selector 305.