Repairs to OXYWAY Fix I, OXYWAY Fix III
70
EN
OXYWAY
Back
11.Unscrew the hand-screwed connection 115 from the bottom part
12.Remove the hand-screwed connection 115 from the counter
13.Remove the connecting bolt 113 from the hand-screwed
14.Remove the rib filter 104 from the connecting bolt 113.
13.8.2 Fitting the connecting bolt
1. Insert the connecting bolt 113 into the hand-screwed
2. Screw the hand-screwed connection 115 onto the bottom part
3. Tightly screw the hand-screwed connection 115 onto the top
4. Insert the rib filter 104 into the connecting bolt 113.
5. Screw the housing onto the connecting bolt 113 by hand.
6. Push the counter cap 404 over the safety valve and the housing.
7. Place the torque wrench onto the square of the counter cap 404.
404
8. Use the torque wrench to tightly screw the housing onto the
113
9. Remove the torque wrench and counter cap 404.
10.Loosen the hand-screwed connection 115 at the top part of the
11.Unscrew the hand-screwed connection 115 from the bottom part
12.Remove the hand-screwed connection 115 of the pressure reducer
13.Remove the top part of the counter tool 400 from the bottom part
14.Loosen the bottom part of the counter tool 400 from the vise and
15.Fit the O-ring 114 in the connecting bolt 113
Forward
of the counter tool 400 by hand.
tool 400.
connection 115.
connection 115.
of the counter tool 400 by hand.
part of the counter tool 400.
connecting bolt 113.
When doing so, note: If using a new connecting bolt 113, it will
need to be loosened with the torque wrench and then re-tightened
again 3 times.
counter tool 400.
of the counter tool 400 by hand.
from the counter tool 400.
of the counter tool 400.
remove.
the O-ring in the connecting bolt", page
Contents
Previous page
(see "13.6.2 Fitting
68).
Next page