Weinmann OXYWAY Fix I Service Manual page 58

Pressure reducer
Table of Contents

Advertisement

Final inspection OXYWAY Click
Accessory
Connection nozzle
Instructions for Use
58
EN
OXYWAY
Back
16.Close the valve on the oxygen cylinder.
17.Remove the special kit 510 from the testing device 510.
18.Remove the special kit 510 from the adapter 507.
19.Remove the adapter 507 from the inhalation outlet 301.
20.Remove the adapter 505 or 506 from the OXYWAY Fix I pressure
21.Remove the adapter 505 or 506 from the O
22.Unscrew the OXYWAY Fix I hand-screwed connection 115 from
11.6 Checking the accessories
1. Also check the accessories according to the following table:
Forward
outlet 105.
the OXYWAY Click.
the valve of the oxygen cylinder.
Requirement
• The connection nozzle shows no signs of external damage.
• The O-ring in the connection nozzle shows no signs of external damage.
The instructions for use are present.