16.11 Replacing the hand-screwed
connection
16.11.1 Removing the hand-screwed connection
Requirement
• The O-ring 272 has been removed from the connecting bolt 271
(see "16.8.1 Removing the O-ring from the connecting bolt",
page
• The connecting bolt 271 has been removed
(see "16.10.1 Removing the connecting bolt", page
1. Dispose of the hand-screwed connection 273.
16.11.2 Fitting the hand-screwed connection
1. Pick up the hand-screwed connection 273.
2. Fit the connecting bolt 271
(see "16.10.2 Fitting the connecting bolt", page
3. Fit the O-ring 272 in the connecting bolt 271
(see "16.8.2 Fitting the O-ring in the connecting bolt", page
16.12 Replacing the rib filter
Required tools
• Watchmaker's screwdriver 409
• Counter tool with fastening handle 400
• Counter cap 404
• Vise
• Torque wrench 60 Nm ± 1 Nm
16.12.1 Removing the rib filter
Requirement
The O-ring 272 has been removed from the connecting bolt 271
(see "16.8.1 Removing the O-ring from the connecting bolt",
page
179).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 273 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 273 onto the top part
of the counter tool 400.
Contents
179).
Previous page
Repairs to OXYWAY Fine I
181).
182).
180).