Weinmann OXYWAY Fix I Service Manual page 197

Pressure reducer
Table of Contents

Advertisement

Back
Forward
5. Fit the spring for the piston 262
(see "16.19.2 Fitting the spring for the piston", page
6. Fit the assembly housing cover 275
(see "16.18.2 Fitting the assembly housing cover", page
7. Fit the driving plate 266
(see "16.17.2 Fitting the driving plate", page
8. Fit the multistep flow selector 267
(see "16.16.2 Fitting the multistep flow selector", page
9. Fit the hexagon nut 268
(see "16.15.2 Fitting the hexagon nut", page
10.Fit the cover for the multistep flow selector 269
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
187).
16.24 Replacing the housing cover
16.24.1 Removing the housing cover
Requirement
• The cover for the multistep flow selector 269 has been removed
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
• The hexagon nut 268 has been removed
(see "16.15.1 Removing the hexagon nut", page
• The multistep flow selector 267 has been removed
(see "16.16.1 Removing the multistep flow selector", page
• The driving plate 266 has been removed
(see "16.17.1 Removing the driving plate", page
• The assembly housing cover 275 has been removed
(see "16.18.1 Removing the assembly housing cover", page
• The spring for the piston 262 has been removed
(see "16.19.1 Removing the spring for the piston", page
• The piston 261 has been removed
(see "16.20.1 Removing the piston", page
• The spring for the lifting pin 263 has been removed
(see "16.21.1 Removing the spring for the lifting pin", page
• The lifting pin with shoulder 265 has been removed
(see "16.22.1 Removing the lifting pin with shoulder", page
• The spindle 270 has been removed
(see "16.23.1 Removing the spindle", page
1. Remove the housing cover 264.
Contents
Repairs to OXYWAY Fine I
189).
188).
192).
196).
OXYWAY
Previous page
192).
190).
188).
187).
188).
188).
189).
191).
193).
194).
EN
197
Next page