Weinmann OXYWAY Fix I Service Manual page 262

Pressure reducer
Table of Contents

Advertisement

OXYWAY Fast I
Pressure reducer with maximum output 15 l/min
x
1 l/min ± 0.26 l/min
x
2 l/min ± 0.36 l/min
x
3 l/min ± 0.5475 l/min
x
4 l/min ± 0.73 l/min
x
5 l/min ± 0.9125 l/min
x
6 l/min ± 1.095 l/min
x
9 l/min ± 1.6425 l/min
x
12 l/min ± 2.19 l/min
x
15 l/min ± 2.7375 l/min
x
Turn from 15 l/min to 6 l/min: 6 l/min ± 1.095 l/min
x
Turn from 6 l/min to 0 l/min: 0 l/min ± 0.04 l/min
OXYWAY Fast I+
Pressure reducer with maximum output 15 l/min
x
0.5 l/min ± 0.11 l/min
x
1 l/min ± 0.26 l/min
x
1.5 l/min ± 0.41 l/min
x
2 l/min ± 0.36 l/min
x
3 l/min ± 0.5475 l/min
x
4 l/min ± 0.73 l/min
x
6 l/min ± 1.095 l/min
x
9 l/min ± 1.6425 l/min
x
15 l/min ± 2.7375 l/min
x
Turn from 15 l/min to 4 l/min: 4 l/min ± 0.73 l/min
x
Turn from 4 l/min to 0 l/min: 0 l/min ± 0.04 l/min
Serial number:
8.8
Checking the accessories
Oxygen cylinder
x
The oxygen cylinder shows no signs of external damage.
x
The valve of the oxygen cylinder is fully functional.
x
The pressure reducer on the oxygen cylinder is fully functional.
x
The oxygen cylinder has a valid test certificate.
Connection nozzle
x
The connection nozzle shows no signs of external damage.
x
The O-ring in the connection nozzle shows no signs of external damage.
The instructions for use are present.
The final inspection was performed and passed.
WEINMANN Emergency Medical Technology GmbH + Co. KG, Frohbösestraße 12, 22525 Hamburg, GERMANY
Test report
According to the service manual WM 66611
Date
Tel: +49 40 88 18 96-120, Fax: +49 40 88 18 96-481
Signature
Page 3 of 3
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
Not
OK
OK
Yes
No