16.6.2 Fitting the sealing disks and nozzle plate in the
inhalation outlet
1. Use tweezers to insert the sealing disk 260 in the inhalation
2. Align the nozzle plate 259 so that the domed part is facing
3. Use tweezers to insert the nozzle plate 259 in the inhalation
4. Use tweezers to insert the sealing disk 260 in the inhalation
5. Fit the pressure screw 258 in the inhalation outlet 257
16.7 Replacing the pressure gauge cap as
protection against impact and falling
16.7.1 Removing the pressure gauge cap as protection
against impact and falling
1. Remove the pressure gauge cap 254 from the contents
16.7.2 Fitting the pressure gauge cap as protection
against impact and falling
1. Place the pressure gauge cap 254 onto the contents indicator 250
16.8 Replacing the O-ring in the
connecting bolt
Required tools
Watchmaker's screwdriver 409
16.8.1 Removing the O-ring from the connecting bolt
1. Use the watchmaker's screwdriver 409 to carefully lift up the
272
Forward
outlet 257.
upwards.
outlet 257.
When doing so, note: The domed part of the nozzle plate 259 must
be pointing towards the opening in the inhalation outlet 257.
outlet 257.
(see "16.5.2 Fitting the pressure screw in the inhalation outlet",
page
178).
indicator 250 or flow indicator 251 by hand.
or flow indicator 251 by hand.
O-ring 272 and remove it from the groove in the connecting
bolt 271.
When doing so, note: The groove must not be damaged.
Contents
Previous page
Repairs to OXYWAY Fine I
OXYWAY
EN
179