Repairs to OXYWAY Fine I
178
EN
OXYWAY
Back
16.5 Replacing the pressure screw in the
inhalation outlet
Required tools
Slot-head screwdriver
16.5.1 Removing the pressure screw from the
inhalation outlet
1. Use the slot-head screwdriver to loosen the pressure screw 258
2. Remove the pressure screw 258 from the inhalation outlet.
257
258
16.5.2 Fitting the pressure screw in the inhalation
outlet
1. Insert the pressure screw 258 into the inhalation outlet 257.
257
258
2. Use the slot-head screwdriver to tightly screw in the pressure
16.6 Replacing the sealing disks and
nozzle plate in the inhalation outlet
Required tools
Tweezers
16.6.1 Removing the sealing disks and nozzle plate
from the inhalation outlet
Requirement
The pressure screw 258 has been removed from the inhalation
outlet 257
inhalation outlet", page
1. Remove the sealing disk 260 from the inhalation outlet 257.
2. Remove the nozzle plate 259 from the inhalation outlet 257.
3. Remove the sealing disk 260 from the inhalation outlet 257.
Forward
in the inhalation outlet 257.
When doing so, note: The slot on the pressure screw 258
must be facing upwards when the screw is fitted.
screw 258 in the inhalation outlet 257.
(see "16.5.1 Removing the pressure screw from the
178).