14.11.2 Fitting the rib filter
1. Insert the rib filter 154 into the connecting bolt 175.
154
175
2. Screw the housing onto the connecting bolt 175 by hand.
3. Push the counter cap 404 over the safety valve and the housing.
404
4. Place the torque wrench onto the square of the counter cap 404.
175
5. Use the torque wrench to tightly screw the housing onto the
connecting bolt 175.
6. Remove the torque wrench and counter cap 404.
7. Loosen the hand-screwed connection 177 at the top part of the
counter tool 400.
8. Unscrew the hand-screwed connection 177 from the bottom part
of the counter tool 400 by hand.
9. Remove the hand-screwed connection 177 of the pressure reducer
from the counter tool 400.
10.Remove the top part of the counter tool 400 from the bottom part
of the counter tool 400.
11.Loosen the bottom part of the counter tool 400 from the vise and
remove.
12.Fit the O-ring 176 in the connecting bolt 175
(see "14.7.2 Fitting the O-ring in the connecting bolt", page
14.12 Replacing the inhalation outlet
Required tools
• Counter tool with fastening handle 400
• Open-end wrench A/F 15
• Vise
• Brass brush
• Loctite 413
• Torque wrench 8 Nm ± 2 Nm
• Thread tap G 1/8
Contents
Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
Previous page
91).