17.23.2 Fitting the holder for the orifice plate
1. Pick up the holder for the orifice plate 313.
2. Assemble the orifice plate assembly 325
(see "17.10.2 Assembling the orifice plate assembly", page
3. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
4. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
17.24 Replacing the orifice plate
17.24.1 Removing the orifice plate
Requirement
• The multistep flow selector assembly 323 has been removed
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
• The multistep flow selector assembly 323 has been disassembled
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
• The orifice plate assembly 325 has been disassembled
(see "17.10.1 Disassembling the orifice plate assembly",
page
215).
1. Dispose of the orifice plate 314.
17.24.2 Fitting the orifice plate
1. Pick up the orifice plate 314.
2. Assemble the orifice plate assembly 325
(see "17.10.2 Assembling the orifice plate assembly", page
3. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
4. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
Contents
Repairs to OXYWAY Click
OXYWAY
Previous page
215).
215).
EN
229