Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
409
94
EN
OXYWAY
Back
14.10.2 Fitting the hand-screwed connection
1. Pick up the hand-screwed connection 177.
2. Fit the connecting bolt 175
3. Fit the O-ring 176 in the connecting bolt 175
14.11 Replacing the rib filter
Required tools
• Watchmaker's screwdriver 409
• Counter tool with fastening handle 400
• Counter cap 404
• Vise
• Torque wrench 60 Nm ± 1 Nm
14.11.1 Removing the rib filter
Requirement
The O-ring 176 has been removed from the connecting bolt 175
(see "14.7.1 Removing the O-ring from the connecting bolt",
page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
3. Screw the hand-screwed connection 177 of the pressure reducer
4. Tightly screw the hand-screwed connection 177 onto the top part
5. Push the counter cap 404 over the safety valve and the housing.
404
6. Place the torque wrench onto the square of the counter cap 404.
175
7. Use the torque wrench to loosen the housing from the connecting
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 175 by hand and
10.Use the watchmaker's screwdriver 409 to remove the rib filter 154
154
175
Forward
(see "14.9.2 Fitting the connecting bolt", page
(see "14.7.2 Fitting the O-ring in the connecting bolt", page
90).
the counter tool 400.
onto the bottom part of the counter tool 400 by hand.
of the counter tool 400.
bolt 175.
then remove.
from the connecting bolt 175.