Weinmann OXYWAY Fix I Service Manual page 43

Pressure reducer
Table of Contents

Advertisement

500
511
Cylinder pressure p1
(in bar)
100
Back
9.7 Checking the constant output (flow)
230
211
Pressure reducer with an output
of Q1:190 l/min
> 160 l/min
Forward
12.Remove the union nut of the supply test set 501 from the pressure
outlet 209.
13.Remove the hand-screwed connection 230 from the valve of the
oxygen cylinder.
at the pressure outlet
1. Screw the hand-screwed connection 230 onto the valve of
the oxygen cylinder.
2. Screw the hand-screwed connection 230 tight by hand.
3. Set the multistep flow selector 211 to "0".
4. Screw the special kit 500 onto the pressure outlet 209.
5. Connect the special kit 500 to the testing device 511.
6. Open the valve of the oxygen cylinder slowly.
7. Read off the cylinder pressure p1 at the contents indicator 200
and enter the value in the test report.
8. Read off the flow values at the testing device 511 and enter
in the test report.
Requirement:
The flow value is within the tolerance range according to the
following table:
Pressure reducer with DIN or
AGA coupling with an output
of Q1:150 l/min
> 120 l/min
9. Close the valve on the oxygen cylinder.
10.Remove the special kit 500 from the testing device 511.
11.Remove the special kit 500 from the pressure outlet 209.
12.Remove the hand-screwed connection 230 from the valve of
the oxygen cylinder.
Contents
Final inspection OXYWAY Fast II
Pressure reducer with
bayonet coupling
> 120 l/min
OXYWAY
Previous page
EN
43
Next page