Weinmann OXYWAY Fix I Service Manual page 270

Pressure reducer
Table of Contents

Advertisement

Serial number:
11.2
Test equipment required
Test equipment (according to service manual OXYWAY pressure reducer WM 66611).
11.3
Preparing the final inspection
x
The pressure reducer shows no signs of external damage.
x
The labels on the pressure reducer are present and legible.
x
The connection threads show no sign of external damage and turn smoothly.
11.4
Checking the system for leaks
x
No bubbles form between the O
x
No bubbles form between the multistep flow selector and the inhalation outlet.
Checking the constant output (flow) at the inhalation outlet
11.5
x
1 l/min ± 0.26 l/min
x
2 l/min ± 0.36 l/min
x
3 l/min ± 0.5475 l/min
x
4 l/min ± 0.73 l/min
x
5 l/min ± 0.9125 l/min
x
6 l/min ± 1.095 l/min
x
9 l/min ± 1.6425 l/min
x
12 l/min ± 2.19 l/min
x
15 l/min ± 2.7375 l/min
x
Turn from 15 l/min to 6 l/min: 6 l/min ± 1.095 l/min
x
Turn from 6 l/min to 0 l/min: 0 l/min ± 0.04 l/min
11.6
Checking the accessories
Connection nozzle
x
The connection nozzle shows no signs of external damage.
x
The O-ring shows no signs of external damage.
The instructions for use are present.
The final inspection was performed and passed.
WEINMANN Emergency Medical Technology GmbH + Co. KG, Frohbösestraße 12, 22525 Hamburg, GERMANY
Test report
According to the service manual WM 66611
sampling point and the housing.
2
Date
Tel: +49 40 88 18 96-120, Fax: +49 40 88 18 96-481
Signature
Page 2 of 2
Not
OK
OK
Not
OK
OK
Not
OK
OK
Not
OK
OK
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
l/min
Not
OK
OK
Yes
No