Weinmann OXYWAY Fix I Service Manual page 137

Pressure reducer
Table of Contents

Advertisement

Back
Forward
13.Remove the top part of the counter tool 400 from the bottom part
of the counter tool 400.
14.Loosen the bottom part of the counter tool 400 from the vise and
remove.
15.Fit the O-ring 229 in the connecting bolt 228
(see "15.8.2 Fitting the O-ring in the connecting bolt", page
15.11 Replacing the hand-screwed
connection
15.11.1 Removing the hand-screwed connection
Requirement
• The O-ring 229 has been removed from the connecting bolt 228
(see "15.8.1 Removing the O-ring from the connecting bolt",
page
• The connecting bolt 228 has been removed
(see "15.10.1 Removing the connecting bolt", page
1. Dispose of the hand-screwed connection 230.
15.11.2 Fitting the hand-screwed connection
1. Pick up the hand-screwed connection 230.
2. Fit the connecting bolt 228
(see "15.10.2 Fitting the connecting bolt", page
3. Fit the O-ring 229 in the connecting bolt 228
(see "15.8.2 Fitting the O-ring in the connecting bolt", page
15.12 Replacing the rib filter
Required tools
• Watchmaker's screwdriver 409
• Counter tool with fastening handle 400
• Counter cap 404
• Vise
• Torque wrench 60 Nm ± 1 Nm
15.12.1 Removing the rib filter
Requirement
The O-ring 229 has been removed from the connecting bolt 228
(see "15.8.1 Removing the O-ring from the connecting bolt",
page
134).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 230 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 230 onto the top part
of the counter tool 400.
Contents
134).
Previous page
Repairs to OXYWAY Fast II
134).
135).
136).
134).
OXYWAY
EN
137
Next page

Advertisement

Table of Contents
loading

Table of Contents