Repairs to OXYWAY Click
415
322
232
EN
OXYWAY
Back
4. Use the special gripping pliers 415 to loosen the housing 320 from
320
323
5. Use the Allen key to loosen the disposable filter 300 from the
300
320
17.27.2 Fitting the housing
1. Use the Allen key to insert the disposable filter 300 into the
300
320
2. Pick up the O
3. Use the open-end wrench to tightly screw in the O
4. Place the multistep flow selector assembly 323 into the cover
5. Tighten the cover clamping device 401 in the vise.
6. Screw the housing 320 with O
320
323
7. Place the 8 Nm torque wrench onto the O
8. Use the 8 Nm torque wrench to tighten the housing 320 onto
9. Slightly loosen the cover clamping device 401.
Forward
the multistep flow selector assembly 323 and then remove it.
housing 320 and then remove it.
housing 320 and then tighten.
sampling point 322 and insert it into the housing 320
2
by hand.
point 322 in the housing 320.
clamping device 401.
When doing so, note: The cover clamping device 401 must clamp
onto the housing cover 308.
multistep flow selector assembly 323 by hand.
the multistep flow selector assembly 323.