Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
176
175
90
EN
OXYWAY
Back
14.6 Replacing the sieve and nozzle plate
in the inhalation outlet
Required tools
• Watchmaker's screwdriver 409
• Tweezers
14.6.1 Removing the sieve and nozzle plate from the
inhalation outlet
Requirement
The pressure screw 156 has been removed from the inhalation
outlet 155
inhalation outlet", page
1. Use the watchmaker's screwdriver 409 to carefully remove the
14.6.2 Fitting the sieve and sealing disks in the
inhalation outlet
1. Use tweezers to insert the sealing disk 158 in the inhalation
2. Use tweezers to insert the sieve 157 in the inhalation outlet 155.
3. Use tweezers to insert the sealing disk 158 in the inhalation
4. Fit the pressure screw 156 in the inhalation outlet 155
14.7 Replacing the O-ring in the
connecting bolt
Required tools
• Watchmaker's screwdriver 409
• Cloth
14.7.1 Removing the O-ring from the connecting bolt
1. Use the watchmaker's screwdriver 409 to carefully lift up the
Forward
(see "14.5.1 Removing the pressure screw from the
89).
sieve 157 and sealing disks 158 from the inhalation outlet 155.
outlet 155.
outlet 155.
(see "14.5.2 Fitting the pressure screw in the inhalation outlet",
page
89).
O-ring 176 and remove it from the groove in the connecting
bolt 175.
When doing so, note: The groove must not be damaged.