4. Place the cover assembly 180 onto the orifice plate assembly 181.
5. Remove the cover assembly 180 from the clamping device for cover
assembly 417 together with the orifice plate assembly 181.
6. Align the orifice plate assembly 181 in the cover assembly 180.
When doing so, note: The bores in the multistep flow selector 161,
the bore in the orifice plate 168, and the grooved pin in the holder
for the orifice plate 169 must be in a straight line.
7. Carefully place the cover assembly 180 and the orifice plate
assembly 181 back onto the clamping device for cover
assembly 417.
When doing so, note: The position of the orifice plate assembly 181
must not change.
8. Insert the needle 169.
9. Insert the sliding ring 164.
10.Insert the springs 165.
11.Place the balls 163 on the springs 165.
12.Place the multistep flow selector 159 on the cover assembly 180
with the needle 169, sliding ring 166, springs 165 and balls 163.
163
When doing so, note: The "0" on the multistep flow selector 159
must be at the level of the needle 169.
165