Repairs to OXYWAY Click
308
220
EN
OXYWAY
Back
17.17 Disassembling/assembling the cover
assembly
Required tools
• Cover clamping device 401
• Vise
• Special counter tool, DISS 416
• Special gripping pliers 415
• Open-end wrench A/F 15
• Watchmaker's screwdriver 409
• Brush
• Oxygen-compatible lubricant 414
• Allen key size 5
• Thread tap G 1/8
• Brass brush
• Torque wrench 8 Nm ± 2 Nm
• Torque wrench 16 Nm ± 1 Nm
• Loctite 413
17.17.1 Disassembling the cover assembly
Requirement
• The multistep flow selector assembly 323 has been removed
• The multistep flow selector assembly 323 has been disassembled
1. Lightly tension the cover clamping device 401 in the vise.
2. Place the cover assembly 324 in the cover clamping device 401.
3. Align the cover assembly 324 in the cover clamping device 401 so
324
401
4. Tighten the cover clamping device 401 in the vise.
Forward
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
that the cover clamping device 401 clamps onto the housing
cover 308.