Repairs to OXYWAY Click
228
EN
OXYWAY
Back
17.22 Replacing the union nut
17.22.1 Removing the union nut
Requirement
• The multistep flow selector assembly 323 has been removed
• The multistep flow selector assembly 323 has been disassembled
• The cover assembly 324 has been disassembled
1. Dispose of the union nut 316.
17.22.2 Fitting the union nut
1. Pick up the union nut 316.
2. Assemble the cover assembly 324
3. Assemble the multistep flow selector assembly 323
4. Fit the multistep flow selector assembly 323
17.23 Replacing the holder for the orifice
plate
17.23.1 Removing the holder for the orifice plate
Requirement
• The multistep flow selector assembly 323 has been removed
• The multistep flow selector assembly 323 has been disassembled
• The orifice plate assembly 325 has been disassembled
1. Dispose of the holder for the orifice plate 167.
Forward
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
(see "17.17.1 Disassembling the cover assembly", page
(see "17.17.2 Assembling the cover assembly", page
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
(see "17.10.1 Disassembling the orifice plate assembly",
page
215).
Contents
Previous page
220).
222).
Next page