15.18 Replacing the threaded locking pin
Required tools
Allen key size 1.5
15.18.1 Removing the threaded locking pin
Requirement
The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
1. Use the Allen key to loosen the threaded locking pin 201
in the housing.
2. Remove the threaded locking pin 201 from the housing.
201
15.18.2 Fitting the threaded locking pin
1. Insert the threaded locking pin 201 into the housing.
201
2. Use the Allen key to lightly screw in the threaded locking pin 201
in the housing.
3. Fit the multistep flow selector assembly 233
(see "15.16.2 Fitting the multistep flow selector assembly",
page
15.19 Replacing the piston
15.19.1 Removing the piston
Requirement
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
• The spring for the piston 227 has been removed
(see "15.17.1 Removing the spring for the piston", page
1. Remove the piston 226 from the multistep flow selector
assembly 233.