Weinmann OXYWAY Fix I Service Manual page 111

Pressure reducer
Table of Contents

Advertisement

Back
Forward
14.25 Replacing the needle
14.25.1 Removing the needle
• The multistep flow selector assembly 179 has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page
• The spring for the piston 174 has been removed
(see "14.14.1 Removing the spring for the piston", page
• The piston 173 has been removed
(see "14.16.1 Removing the piston", page
• The multistep flow selector assembly 179 has been disassembled
(see "14.17.1 Disassembling the multistep flow selector assembly",
page
1. Dispose of the needle 169.
169
14.25.2 Fitting the needle
1. Pick up the needle 169.
2. Assemble the multistep flow selector assembly 179
(see "14.17.2 Assembling the multistep flow selector assembly",
page
3. Fit the piston 173
4. Fit the spring for the piston 174
(see "14.14.2 Fitting the spring for the piston", page
5. Fit the multistep flow selector assembly 179
(see "14.13.2 Fitting the multistep flow selector assembly",
page
14.26 Replacing the sliding ring
14.26.1 Removing the sliding ring
• The multistep flow selector assembly 179 has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page
• The spring for the piston 174 has been removed
(see "14.14.1 Removing the spring for the piston", page
• The piston 173 has been removed
(see "14.16.1 Removing the piston", page
Contents
Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
98).
101).
102).
(see "14.16.2 Fitting the piston", page
98).
98).
Previous page
99).
100).
101).
99).
99).
100).
OXYWAY
EN
111
Next page