Weinmann OXYWAY Fix I Service Manual page 190

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fine I
404
402
404
190
EN
OXYWAY
Back
1. Lightly tension the cover clamping device 401 in the vise.
2. Insert the pressure reducer into the cover clamping device 401.
3. Tighten the cover clamping device 401 in the vise.
4. Push the counter cap 404 over the safety valve and the housing.
5. Place the torque wrench onto the square of the counter cap 404.
6. Use the torque wrench to slightly loosen the housing from the
264
7. Remove the torque wrench and counter cap 404.
8. Loosen the vise and carefully remove the pressure reducer from the
9. Carefully loosen the assembly housing cover 275 from the housing
16.18.2 Fitting the assembly housing cover
1. Insert the pressure reducer housing into the cover screw-in
2. Insert the assembly housing cover 275 into the housing.
3. Press down the lever on the cover screw-in device 402 until the
275
4. Maintain the tension and screw the assembly housing cover 275
5. Carefully release the tension on the cover screw-in device 402.
6. Remove the pressure reducer from the cover screw-in device 402.
7. Insert the pressure reducer into the cover clamping device 401.
8. Tighten the cover clamping device 401 in the vise.
9. Push the counter cap 404 over the safety valve and the housing.
10.Place the torque wrench onto the square of the counter cap 404.
11.Use the torque wrench to tightly screw the housing onto the
275
Forward
When doing so, note: The cover clamping device 401 clamps onto
the housing cover 264.
housing cover 264.
When doing so, note: The housing must not yet be removed.
cover clamping device 401.
by hand and then remove.
device 402.
assembly housing cover 275 is resting gently on the housing.
onto the housing by hand.
When doing so, note: The cover clamping device 401 clamps onto
the assembly housing cover 275.
assembly housing cover 275.