Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
181
162
181
417
102
EN
OXYWAY
Back
The balls and sliding ring on the bottom of the multistep flow selector
may stick.
3. Remove the balls 163.
4. Remove the springs 165.
5. Remove the sliding ring 164.
6. Remove the needle 169.
7. Use the Allen key to push the orifice plate assembly 181 into
8. Remove the orifice plate assembly 181.
9. If necessary: Disassemble the orifice plate assembly 181
10.If necessary: Disassemble the cover assembly 180
14.17.2 Assembling the multistep flow selector
assembly
1. If necessary: Assemble the cover assembly 180
2. If necessary: Assemble the orifice plate assembly 181
3. Pick up the orifice plate assembly 181 and place it onto the
Forward
the inside of the housing cover 162.
(see "14.20.1 Disassembling the orifice plate assembly",
page
106).
(see "14.27.1 Disassembling the cover assembly", page
(see "14.27.2 Assembling the cover assembly", page
(see "14.20.2 Assembling the orifice plate assembly", page
clamping device for cover assembly 417.