Repairs to OXYWAY Click
415
316
322
224
EN
OXYWAY
Back
22.Turn the multistep flow selector 307 with inhalation outlet 301
23.Place the special gripping pliers 415 on the union nut 316 and hold
301
24.Place the 16 Nm torque wrench onto the inhalation outlet 301.
25.Tighten the multistep flow selector 307 with inhalation outlet 301
307
26.Turn the housing 320 with the cover assembly 324 another 180°
27.Place the torque wrench onto the hexagon of the O
28.Use the 16 Nm torque wrench to loosen the housing 320 from
29.Remove the housing 320 with O
320
324
30.Loosen the vise and remove the cover assembly 324 from the cover
31.Remove the cover clamping device 401 from the vise.
32.Assemble the multistep flow selector assembly 323
33.Fit the multistep flow selector assembly 323
Forward
around 30° counter-clockwise by hand in order to reach the desired
position after tightly screwing on.
in place.
using the special gripping pliers 415 and the 16 Nm torque wrench.
When doing so, note: The torque wrench must activate in the
desired position at 16+ 2 Nm.
and place it into the cover clamping device 401.
When doing so, note: The cover clamping device 401 must clamp
onto the housing cover 308.
point 322.
the cover assembly 324.
cover assembly 324 by hand.
clamping device 401.
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).