Weinmann OXYWAY Fix I Service Manual page 226

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Click
226
EN
OXYWAY
Back
17.19 Replacing the O-rings in the
multistep flow selector
17.19.1 Removing the O-rings from the multistep flow
selector
Requirement
• The multistep flow selector assembly 323 has been removed
• The multistep flow selector assembly 323 has been disassembled
• The cover assembly 324 has been disassembled
1. Dispose of the large O-ring 318.
2. Dispose of the small O-ring 317.
17.19.2 Fitting the O-rings in the multistep flow selector
1. Insert the large O-ring 318 into the groove in the multistep flow
2. Insert the small O-ring 317 into the groove in the multistep flow
3. Assemble the cover assembly 324
4. Assemble the multistep flow selector assembly 323
5. Fit the multistep flow selector assembly 323
17.20 Replacing the housing cover
17.20.1 Removing the housing cover
Requirement
• The multistep flow selector assembly 323 has been removed
• The multistep flow selector assembly 323 has been disassembled
• The cover assembly 324 has been disassembled
• The O-ring 319 has been removed from the housing cover 308
1. Dispose of the housing cover 308.
Forward
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
(see "17.17.1 Disassembling the cover assembly", page
selector 307 by hand.
selector 307 by hand.
(see "17.17.2 Assembling the cover assembly", page
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
(see "17.17.1 Disassembling the cover assembly", page
(see "17.18.1 Removing the O-ring from the housing cover",
page
225).
Contents
Previous page
220).
222).
220).
Next page