Final inspection OXYWAY Click
56
EN
OXYWAY
Back
11.4 Checking the system for leaks
1. Screw the OXYWAY Fix I hand-screwed connection 115 onto
2. Screw the OXYWAY Fix I hand-screwed connection 115 tight
3. Set the OXYWAY Click multistep flow selector 305 to "0".
4. Connect the adapter 505 or 506 to the OXYWAY Fix I pressure
5. Connect the other end of the adapter 505 or 506 to the O
6. Open the valve of the oxygen cylinder slowly.
7. Use a brush to lightly moisten the connection between the O
8. Screw the cap 504 onto the inhalation outlet 301.
9. Set the multistep flow selector 305 to "15".
10.Use a brush to lightly moisten the connection between the
11.Set the multistep flow selector 305 to "0".
12.Remove the cap 504 from the inhalation outlet 301.
13.Close the valve on the oxygen cylinder.
14.Remove the adapter 505 or 506 from the O
15.Remove the adapter 505 or 506 from the OXYWAY Fix I pressure
16.Unscrew the OXYWAY Fix I hand-screwed connection 115 from
17.Wipe the OXYWAY Click with a clean, dry cloth.
Forward
the valve of the oxygen cylinder.
by hand.
outlet 105.
sampling point on the OXYWAY Click.
sampling point and the housing 320 with leak detection spray 502.
Requirement:
No bubbles form between the O
multistep flow selector 307 and the inhalation outlet 301 with
leak detection spray 502.
Requirement:
No bubbles form between the multistep flow selector 307 and
the inhalation outlet 301.
the OXYWAY Click.
outlet 105.
the valve of the oxygen cylinder.