Weinmann OXYWAY Fix I Service Manual page 69

Pressure reducer
Table of Contents

Advertisement

113
103
Back
13.7.2 Fitting the disposable filter
1. Insert the disposable filter 103 into the connecting bolt 113.
2. Use the Allen key to tightly screw the disposable filter 103 in
13.8 Replacing the connecting bolt
Required tools
• Counter tool with fastening handle 400
• Vise
• Counter cap 404
• Torque wrench 60 Nm ± 1 Nm
13.8.1 Removing the connecting bolt
Requirement
The O-ring 114 has been removed from the connecting bolt 113
(see "13.6.1 Removing the O-ring from the connecting bolt",
page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
3. Screw the hand-screwed connection 115 of the pressure reducer
4. Tightly screw the hand-screwed connection 115 onto the top part
5. Push the counter cap 404 over the safety valve and the housing.
6. Place the torque wrench onto the square of the counter cap 404.
404
7. Use the torque wrench to loosen the housing from the connecting
113
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 113 by hand
10.Loosen the hand-screwed connection 115 at the top part of
Forward
the connecting bolt 113.
68).
the counter tool 400.
onto the bottom part of the counter tool 400 by hand.
of the counter tool 400.
bolt 113.
and then remove.
the counter tool 400.