15.27 Replacing the springs
15.27.1 Removing the springs
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
• The spring for the piston 227 has been removed
(see "15.17.1 Removing the spring for the piston", page
• The piston 226 has been removed
(see "15.19.1 Removing the piston", page
• The multistep flow selector assembly 233 has been disassembled
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
148).
1. Dispose of the springs 217.
15.27.2 Fitting the springs
1. Pick up the springs 217.
2. Assemble the multistep flow selector assembly 233
(see "15.20.2 Assembling the multistep flow selector assembly",
page
149).
3. Fit the piston 226
4. Fit the spring for the piston 227
(see "15.17.2 Fitting the spring for the piston", page
5. Fit the multistep flow selector assembly 233
(see "15.16.2 Fitting the multistep flow selector assembly",
page
145).
15.28 Replacing the needle
15.28.1 Removing the needle
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
• The spring for the piston 227 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
• The piston 226 has been removed
(see "15.19.1 Removing the piston", page
• The multistep flow selector assembly 233 has been disassembled
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
148).
Contents
Repairs to OXYWAY Fast II
147).
(see "15.19.2 Fitting the piston", page
147).
OXYWAY
Previous page
146).
148).
146).
EN
157
Next page