Repairs to OXYWAY Fast II
158
EN
OXYWAY
Back
1. Dispose of the needle 221.
221
15.28.2 Fitting the needle
1. Pick up the needle 221.
2. Assemble the multistep flow selector assembly 233
3. Fit the piston 226
4. Fit the spring for the piston 227
5. Fit the multistep flow selector assembly 233
15.29 Replacing the sliding ring
15.29.1 Removing the sliding ring
• The multistep flow selector assembly 233 has been removed
• The spring for the piston 227 has been removed
• The piston 226 has been removed
• The multistep flow selector assembly 233 has been disassembled
1. Dispose of the sliding ring 216.
15.29.2 Fitting the sliding ring
1. Pick up the sliding ring 216.
2. Assemble the multistep flow selector assembly 233
3. Fit the piston 226
4. Fit the spring for the piston 227
5. Fit the multistep flow selector assembly 233
Forward
(see "15.20.2 Assembling the multistep flow selector assembly",
page
149).
(see "15.19.2 Fitting the piston", page
(see "15.17.2 Fitting the spring for the piston", page
(see "15.16.2 Fitting the multistep flow selector assembly",
page
145).
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
(see "15.17.1 Removing the spring for the piston", page
(see "15.19.1 Removing the piston", page
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
148).
(see "15.20.2 Assembling the multistep flow selector assembly",
page
149).
(see "15.19.2 Fitting the piston", page
(see "15.17.2 Fitting the spring for the piston", page
the multistep flow selector assembly", page
Contents
Previous page
148).
146).
146).
147).
148).
146).
(see "15.16.2 Fitting
145).
Next page

Advertisement

Table of Contents
loading

Table of Contents