Weinmann OXYWAY Fix I Service Manual page 92

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
92
EN
OXYWAY
Back
14.9 Replacing the connecting bolt
Required tools
• Counter tool with fastening handle 400
• Counter cap 404
• Vise
• Torque wrench 60 Nm ± 1 Nm
14.9.1 Removing the connecting bolt
Requirement
The O-ring 176 has been removed from the connecting bolt 175
(see "14.7.1 Removing the O-ring from the connecting bolt",
page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of the
3. Screw the hand-screwed connection 177 of the pressure reducer
4. Tightly screw the hand-screwed connection 177 onto the top part
5. Push the counter cap 404 over the safety valve and the housing.
404
6. Place the torque wrench onto the square of the counter cap 404.
175
7. Use the torque wrench to loosen the housing from the connecting
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 175 by hand and
10.Loosen the hand-screwed connection 177 at the top part of the
11.Unscrew the hand-screwed connection 177 from the bottom part
12.Remove the hand-screwed connection 177 from the counter
13.Remove the connecting bolt 175 from the hand-screwed
14.Remove the rib filter 154 from the connecting bolt 175.
Forward
90).
counter tool 400.
onto the bottom part of the counter tool 400 by hand.
of the counter tool 400.
bolt 175.
then remove.
counter tool 400.
of the counter tool 400 by hand.
tool 400.
connection 177.
Contents
Previous page
Next page

Advertisement

Table of Contents
loading

Table of Contents