Vise - Weinmann OXYWAY Fix I Service Manual

Pressure reducer
Table of Contents

Advertisement

203
228
228
203
Back
Forward
15.9 Replacing the disposable filter
Required tools
Allen key size 5
15.9.1 Removing the disposable filter
1. Use the Allen key to loosen the disposable filter 203 in the
connecting bolt 228.
2. Remove the disposable filter 203 from the connecting bolt 228.
15.9.2 Fitting the disposable filter
1. Insert the disposable filter 203 into the connecting bolt 228.
2. Use the Allen key to tightly screw the disposable filter 203
in the connecting bolt 228.
15.10 Replacing the connecting bolt
Required tools
• Counter tool with fastening handle 400
• Counter cap 404

• Vise

• Torque wrench 60 Nm ± 1 Nm
15.10.1 Removing the connecting bolt
Requirement
The O-ring 229 has been removed from the connecting bolt 228
(see "15.8.1 Removing the O-ring from the connecting bolt",
page
134).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 230 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
Contents
Repairs to OXYWAY Fast II
OXYWAY
Previous page
EN
135
Next page

Advertisement