Repairs to OXYWAY Fast II
144
EN
OXYWAY
Back
5. Insert the O
232
6. Use the torque wrench to tightly screw in the O
7. Loosen the hand-screwed connection 230 at the top part of the
8. Unscrew the hand-screwed connection 230 from the bottom part
9. Remove the hand-screwed connection 230 of the pressure reducer
10.Remove the top part of the counter tool 400 from the bottom part
11.Loosen the bottom part of the counter tool 400 from the vise and
12.Fit the O-ring 229 in the connecting bolt 228
15.16 Replacing the multistep flow selector
assembly
Required tools
• Cover clamping device 401
• Cover screw-in device 402
• Counter cap 404
• Torque wrench 8 Nm ± 2 Nm
• Vise
• Pressure piece for the cover screw-in device 403
15.16.1 Removing the multistep flow selector assembly
1. Lightly tension the cover clamping device 401 in the vise.
2. Insert the pressure reducer into the cover clamping device 401.
3. Tighten the cover clamping device 401 in the vise.
Forward
coupling 232 and reducer into the housing.
2
and reducer in the housing.
When doing so, note: Loctite 413 must harden according to the
manufacturer's specifications.
counter tool 400.
of the counter tool 400 by hand.
from the counter tool 400.
of the counter tool 400.
remove.
(see "15.8.2 Fitting the O-ring in the connecting bolt", page
When doing so, note: The cover clamping device 401 must
clamp onto the housing cover 214.