Weinmann OXYWAY Fix I Service Manual page 79

Pressure reducer
Table of Contents

Advertisement

Back
Forward
13.15 Replacing the threaded locking pin
Required tools
Allen key size 1.5
13.15.1 Removing the threaded locking pin
Requirement
The spring for the piston 111 has been removed
(see "13.13.1 Removing the spring for the piston", page 77)
1. Use the Allen key to loosen the threaded locking pin 101 in
the housing.
2. Remove the threaded locking pin 101 from the housing.
101
13.15.2 Fitting the threaded locking pin
1. Insert the threaded locking pin 101 into the housing.
101
2. Use the Allen key to lightly screw in the threaded locking pin 101.
3. Fit the spring for the piston 111
(see "13.13.2 Fitting the spring for the piston", page
13.16 Replacing the contents indicator
Required tools
• Oxygen-compatible lubricant 414
• Brush
13.16.1 Removing the contents indicator
Requirement
• The spring for the piston 111 has been removed
(see "13.13.1 Removing the spring for the piston", page 77)
• The threaded locking pin 101 has been removed
(see "13.15.1 Removing the threaded locking pin", page 79)
Contents
Repairs to OXYWAY Fix I, OXYWAY Fix III
OXYWAY
Previous page
77).
EN
79
Next page

Advertisement

Table of Contents
loading

Table of Contents