Repairs to OXYWAY Fix I, OXYWAY Fix III
74
EN
OXYWAY
Back
6. Remove the pressure outlet 105/inhalation outlet 106 from
105/106
7. Remove any adhesive residue from the threads on the pressure
13.11.2 Fitting the pressure outlet/inhalation outlet
1. Pick up the pressure outlet 105/inhalation outlet 106.
Risk due to adhesive in the gas flow!
If the first two threads of the pressure outlet/inhalation outlet are
moistened with adhesive, this adhesive can get into the gas flow to
the patient and lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the pressure outlet 105/
3. Insert the pressure outlet 105/inhalation outlet 106 into the
105/106
4. Use the torque wrench to tightly screw the pressure outlet 105/
5. Loosen the hand-screwed connection 115 at the top part of
6. Unscrew the hand-screwed connection 115 from the bottom
7. Remove the hand-screwed connection 115 of the pressure
8. Remove the top part of the counter tool 400 from the bottom
9. Loosen the bottom part of the counter tool 400 from the vise
Forward
the housing.
reducer housing.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
inhalation outlet 106.
housing.
inhalation outlet 106 in the housing.
When doing so, note: Loctite 413 must harden according to
the manufacturer's specifications.
the counter tool 400.
part of the counter tool 400 by hand.
reducer from the counter tool 400.
part of the counter tool 400.
and remove.