Weinmann OXYWAY Fix I Service Manual page 182

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fine I
404
182
EN
OXYWAY
Back
16.10.2 Fitting the connecting bolt
1. Pick up the connecting bolt 271.
2. Insert the connecting bolt 271 into the hand-screwed
3. Screw the hand-screwed connection 273 onto the bottom part of
4. Tightly screw the hand-screwed connection 273 onto the top part
5. Insert the rib filter 256 into the connecting bolt 271.
6. Screw the housing onto the connecting bolt 271 by hand.
7. Push the counter cap 404 over the safety valve and the housing.
8. Place the torque wrench onto the square of the counter cap 404.
271
9. Use the torque wrench to tightly screw the housing onto the
10.Remove the torque wrench and counter cap 404.
11.Loosen the hand-screwed connection 273 at the top part of the
12.Unscrew the hand-screwed connection 273 from the bottom part
13.Remove the hand-screwed connection 273 of the pressure reducer
14.Remove the top part of the counter tool 400 from the bottom part
15.Loosen the bottom part of the counter tool 400 from the vise and
16.Fit the O-ring 272 in the connecting bolt 271
Forward
connection 273.
the counter tool 400 by hand.
of the counter tool 400.
connecting bolt 271.
When doing so, note: If using a new connecting bolt 271, it will
need to be loosened with the torque wrench and then re-tightened
again 3 times.
counter tool 400.
of the counter tool 400 by hand and then remove.
from the counter tool 400.
of the counter tool 400.
remove.
(see "16.8.2 Fitting the O-ring in the connecting bolt", page
Contents
Previous page
180).
Next page