Weinmann OXYWAY Fix I Service Manual page 149

Pressure reducer
Table of Contents

Advertisement

235
417
Back
The balls and sliding ring on the bottom of the multistep flow selector
may stick.
3. Remove the balls 215.
4. Remove the springs 217.
5. Remove the sliding ring 216.
6. Remove the needle 221.
7. Use the Allen key to push the orifice plate assembly 235 into the
235
214
8. Remove the orifice plate assembly 235.
9. If necessary: Disassemble the orifice plate assembly 235
10.If necessary: Disassemble the cover assembly 234
15.20.2 Assembling the multistep flow selector
assembly
1. If necessary: Assemble the cover assembly 234
2. If necessary: Assemble the orifice plate assembly 235
3. Pick up the orifice plate assembly 235 and place it onto the
Forward
inside of the housing cover 214.
(see "15.23.1 Disassembling the orifice plate assembly",
page
153).
(see "15.30.1 Disassembling the cover assembly", page
(see "15.30.2 Assembling the cover assembly", page
(see "15.23.2 Assembling the orifice plate assembly", page
clamping device for cover assembly 417.
Contents
Previous page
Repairs to OXYWAY Fast II
159).
161).
153).
OXYWAY
EN
149
Next page