10.If an inhalation outlet 106 is present: Fit the sealing disk 109
and nozzle plate 107 in the inhalation outlet 106
(see "13.5.2 Fitting the sealing disks and nozzle plate in the
inhalation outlet", page
11.Fit the pressure screw 108 into the pressure outlet 105/inhalation
outlet 106
outlet/inhalation outlet", page
12.Fit the O-ring 114 in the connecting bolt 113
(see "13.6.2 Fitting the O-ring in the connecting bolt", page
13.12 Replacing the O
Required tools
• Counter tool with fastening handle 400
• Vise
• Loctite 413
• Open-end wrench A/F 19
• Torque wrench 16 Nm ± 1 Nm
• Brass brush
• Thread tap G 1/8
13.12.1 Removing the O
Requirement
The O-ring 114 has been removed from the connecting bolt 113
(see "13.6.1 Removing the O-ring from the connecting bolt",
page
68).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 115 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 115 onto the top
part of the counter tool 400.
5. Use the open-end wrench to loosen the O
reducer from the pressure reducer and then remove it.
117
6. Remove any adhesive residue from the threads on the pressure
reducer housing. When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.