14.20.2 Assembling the orifice plate assembly
1. Align the orifice plate 168 in your hand.
When doing so, note: The largest bore on the orifice plate 168
must be on the left-hand side.
2. Use a brush to apply a thin coat of oxygen-compatible lubricant 414
to the side of the orifice plate 168 that is facing upwards.
3. Place the orifice plate 168 on the holder for the orifice plate 167.
When doing so, note: The grooved pin in the holder for the orifice
plate 167 must be inside the hole in the orifice plate 168.
4. Pick up the slide disk 166.
5. Align the slide disk 166 in your hand.
When doing so, note: The side of the slide disk 166 that has
markings indicating another hole in addition to the hole for
166
the grooved pin must be facing upwards.
6. Use a brush to apply a thin coat of oxygen-compatible lubricant 414
to the side of the slide disk 166 that is facing upwards.
7. Place the slide disk 166 on the holder for the orifice plate 167 and
166
the orifice plate 168.
When doing so, note: The grooved pin in the holder for the orifice
168
plate 167 must be inside the hole in the slide disk 166.
167
8. Assemble the multistep flow selector assembly 179
(see "14.17.2 Assembling the multistep flow selector assembly",
page
102).
9. Fit the piston 173
10.Fit the spring for the piston 174
(see "14.14.2 Fitting the spring for the piston", page
11.Fit the multistep flow selector assembly 179
(see "14.13.2 Fitting the multistep flow selector assembly",
page
98).